| Test Voltage |
• DC : 0 to 5000 V continuously variable by multi turn potentiometer
AC : 0 to 5000 V continuously variable by multi turn potentiometer (with AC or DC test voltage selector switch) |
|
|
Tripping Current
AC and DC |
: adjustable 1.0 to 30.0 mA by press to set switch and
potentiometer on mA meter |
|
|
Accuracy class of Test Voltage |
: ± 0.1%± 2 least count
Resolution: 4½ digit |
|
|
| Accuracy of leakage Current |
: ± 0.5% ± 2 least count
Resolution: 3½ digit |
|
|
| Display Digital |
: Voltage: 4½ digit , Current: 3½ digit |
|
|
| Automatic Timer |
: 1 to 999 second Settable by thumbwheel switches on 7 segment 3 digit display With timer ON/OFF switch |
|
|
| Indication for Pass/Fail |
: Audio (Buzzer) and visual (Neon) alarms when HV is ON,
Output Trips when breakdown occurs i.e. leakage exceeds than set current. Test FAIL indicated by LED
If device under test successfully handles the test voltage for the set time duration then Test PASS indicated by LED |
|
|
| Power Supply |
: 230V±10%, single phase, 50Hz |
|
|